12 Specifications

CPR Targets

Category

Specification

Compression Depth Target

> 50 mm
( 2”)
Depth accuracy: ±10%

Compression Release Target
Force

< 2,5 kg (5,5 lbs).
Force accuracy: +1,5 kg, -2,0 kg
(+3,3 lbs, - 4,4 lbs)

Compression Rate Target

100 to 120/min ± 3/min

CPRmeter [REF 801-00140]

The CPRmeter meets the performance requirements of IEC 60601-1, 2nd
and 3rd edition.

Category

Specification

Dimensions

154 mm x 64 mm x 28 mm
(6.1” x 2.5” x 1.1”)

Weight

227g (7,3 oz)

Battery

3V Lithium 123 (Li/MnO

2

)

The CPRmeter has been tested
with Energizer

®

batteries. Use

only Energizer

®

, Panasonic

®

or

Duracell

®

batteries.

microSD™ Memory Card

The CPRmeter has been tested
with 2Gb microSD™ memory
cards. Always test a new card
before using it in data collection.

Temperature and Relative
Humidity

Storage temperature:
-20° to 70° C (-4° to 158° F).
Relative humidity: 0% to 75%.
Operating temperature:
0° to 40° C (32° to 104° F).
Operating atmospheric pressure :
1000 to 540 mbar (100 to 54 kPa).
Relative humidity 0% to 95%.
Sealing: Meets ISO/IEC 60529 class
IP55.
Product temperature can reach
60° C.

Electromagnetic Compatiblity

Meets IEC 60601-1-2 and RTCA/
DO-160E.

Recovery time

Recovery time after defibrillation: 0 s