Measuring capacitance, 21 measuring capacitance, Multimeters making measurements – Fluke 87 V User Manual

Page 29: Figure 6. measuring capacitance, Select capacitance

Advertising
background image

Multimeters

Making Measurements

21

Measuring Capacitance

W

Caution

To avoid possible damage to the Meter or to
the equipment under test, disconnect circuit
power and discharge all high-voltage
capacitors before measuring capacitance.
Use the dc voltage function to confirm that
the capacitor is discharged.

The Meter's capacitance ranges are 10.00 nF, 100.0 nF,
1.000

μ

F, 10.00

μ

F, 100.0

μ

F, and 9999

μ

F.

To measure capacitance, set up the Meter as shown in
Figure 6.
To improve the accuracy of measurements less than
1000 nF, use the relative (REL) mode to subtract the
residual capacitance of the Meter and leads.

Note

If too much electrical charge is present on the
capacitor being tested, the display shows
diSC".

MIN MAX

RANGE

REL

Hz %

AutoHOLD

Peak MIN MAX

4½ DIGITS

1 Second

˚

C/

˚

F

OFF

mA

A

mV

V

V

A

A

mA

COM

V

400mA
FUSED

10A MAX

FUSED

A

TRUE RMS MULTIMETER

87 V

LO

LO

Select

Capacitance

+

+
+

+

+

+
+

+

+

aom10f.eps

Figure 6. Measuring Capacitance