Fluke 87 V User Manual

Page 60

Advertising
background image

80 Series V
Users Manual

52

Table 19. MIN MAX Recording Specifications

Model Nominal

Response

Accuracy

83

100 ms to 80 %

Specified accuracy

±

12 counts for changes > 200 ms in duration

(

±

40 counts in ac with beeper on)

100 ms to 80 % (dc functions)

Specified accuracy

±

12 counts for changes > 200 ms in duration

120 ms to 80 % (ac functions)

Specified accuracy

±

40 counts for changes > 350 ms and inputs > 25 % of range

87

250

μ

s (peak) (Model 87 only)

1

Specified

accuracy

±

100 counts for changes > 250

μ

s in duration

(add

±

100 counts for readings over 6000 counts)

(add

±

100 counts for readings in Low Pass mode)

1. For repetitive peaks: 1 ms for single events.