Specifications – Fluke 87 V User Manual

Page 51

Advertising
background image

Multimeters

Specifications

43

Specifications

Maximum Voltage between any Terminal and Earth Ground: 1000 V rms

W

Fuse Protection for mA or

μ

A inputs: 44/100 A, 1000 V FAST Fuse

W

Fuse Protection for A input: 11 A, 1000 V FAST Fuse

Display: Digital: 6000 counts updates 4/sec; (Model 87 also has 19,999 counts in high-resolution mode).
Analog Bargraph: 33 segments, updates 40/sec. Frequency: 19,999 counts, updates 3/sec at >10 Hz
Temperature:
Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C
Altitude:
Operating: 2000 m; Storage: 10,000 m
Temperature Coefficient:
0.05 x (specified accuracy)/ °C (<18 °C or >28 °C)
Electromagnetic Compatibility:

In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts

Except: 600

μ

A dc range total accuracy=specified accuracy +60 counts.

Temperature not specified.

Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 55 °C)
Battery Type:
9 V zinc, NEDA 1604 or 6F22 or 006P
Battery Life:
400 hrs typical with alkaline (with backlight off)
Vibration:
Per MIL-PRF-28800 for a Class 2 instrument
Shock:
1 Meter drop per IEC 61010-1:2001
Size (HxWxL):
1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with Holster and Flex-Stand:
2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight:
12.5 oz (355 g)
Weight with Holster and Flex-Stand:
22.0 oz (624 g)
Safety:
Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC 664 to 600 V
Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1.
IP Rating: 30