Dynex OPSYS MW 91000051 User Manual

Page 50

Advertising
background image

Creating a New Wash Protocol

To create a new wash protocol header:

1. Display the main menu.

1: - - - - - - - - - -
SETUP PROGRAM

2. Use the up or down scroll keys to display the protocol number (or

use the keypad to enter the protocol number) that is to be
created:

6: - - - - - - - - - -
SETUP PROGRAM

Note: Dashes will be displayed if a protocol
has not yet been named.

3. Select PROGRAM. The name of the wash protocol is displayed

in the lower line:

Enter name
- - - - - - - - - -

4. Use the keypad to enter the protocol name (up to 9 characters)

and press ENTER. The display will read:

Password?
YES NO

Note: Refer to page 24 for instructions on
entering alphanumeric characters.

5. Select YES if the protocol is to be password protected. The

display will read:

Enter password
********

Note: If NO is selected, the Select plate
prompt (Step 7) will be displayed.

6. Enter the password, (up to eight numeric characters) and press

ENTER. The display will read:

Confirm password
********

42

Opsys MW™ User’s Guide