Dynex OPSYS MW 91000051 User Manual

Page 88

Advertising
background image

Operational Problems

Symptom

Probable Cause Resolution

Aspiration is not
complete.
(continued)

Hydrophobic filter
may be
obstructed.

Replace filter.

High Waste Level
is displayed.

The Waste
Container is full.

Empty the Waste
Container (see page 70).

The waste level
sensor cable is
not plugged in.

Connect the waste level
sensor cable (see
page 17).

The waste level
sensor connector
is defective.

Service is required.

The waste level
sensor detector is
reversed.

Service is required.

The waste
container is
deformed.

The waste tubing
is blocked.

Remove the waste tubing
from the rear of the
instrument (see page 17).
Check the waste tubing
(and aspiration tubing)
and remove any
blockages or kinks.

The waste
container is
damaged or
defective.

Replace the waste
container.

X Motor Home
Error
is displayed.

Motion of the
wash head is
being blocked.

Check the wash head
vicinity and remove any
obstacles.

The motor is
defective.

Service is required.

Z Motor Home
Error
is displayed.

Motion of the
wash head is
being blocked.

Check the wash head
vicinity and remove any
obstacles.

The motor is
defective.

Service is required.

Test Too Large is
displayed.

The wash
protocol exceeds
the memory limit.

Reduce the number of
steps in the wash protocol
and try again.

80

Opsys MW™ User’s Guide

Advertising