Replacing the Tubing

To replace the vacuum tubing:

Note: This procedure describes replacement
of the vacuum tubing connecting the Waste
Container to the Foam/Aerosol Trap. The
vacuum tubing connecting the Foam/Aerosol
Trap to the rear of the instrument should not
require changing.

1. Unscrew the Cap of the Waste Container, lay it and the level

sensor assembly on a clean paper towel.

2. Discard the contents of the Waste Container in accordance with

local regulations. Thoroughly rinse the container with deionized
water and let it dry.

3. Disconnect the vacuum tubing from the connector in the Cap and

from the connector in the Foam/Aerosol Trap. Discard the old
tube.

4. Obtain a new length of vacuum tubing.

5. Connect one end of the vacuum tubing to the connector in the

Waste Container Cap and the other end to the connector in the
Foam/Aerosol Trap.

6. Place the level sensor assembly back into the Waste Container

and screw the Cap on.

72

Opsys MW™ User’s Guide