Pump

Calibration

8. If the dispensed volume is too high, decrease the pump speed

using the SETUP/CAL function (refer to page 36).

• A pump speed change of -10 units roughly corresponds to a

well volume change of -10µl.

9. If the dispensed volume is too low, increase the pump speed

using the SETUP/CAL function (refer to page 36).

• A pump speed change of +10 units roughly corresponds to a

well volume change of +10µl.

10. Repeat Steps 3-9 until fluid is dispensed with the accuracy

required. (e.g., Error less than or equal to ±5%)