9.1. Test Tasks - Layer 1: DUT-OSC

DApps-MPT will start examining Crystal Oscill

ator’s frequency of the DUT and see if it’s either faster or

slower than standard speed in ppm scale.

A. Setup

Test Port: Select the port that will be used for DUT-OSC test.

Media Type: Click the Media Type scroll-

down menu to choose DUT’s link speed.

Minimum Waiting Time: This task will halt at least for the Minimum Waiting Time you input here

during auto-negotiation process.

Media Type Waiting Timeout: If the time spent for auto-negotiation exceeds the Media Type

Waiting Timeout you set here, the test will stop and the test result will be fail.

B. Criteria

OSC PPM Min/Max: Please input the minimum/maximum frequencies (ppm) which will serve as

DUT-

OSC Task’s testing criteria.

C. Misc

Title of task name: You can assign a name to this task for identification.

Apply: Apply the changes you made.

D. Help

All test variables used for this task and their definitions will be listed here for reference.