Iq information, Hardware settings – Dukane Ultrasonic Hand Held System iQ Series User Manual

Page 33

Advertising
background image

Page  7

Section 5 - Process Control Settings

Dukane Manual Part No. 403-577-01

System Information,

Hardware Settings, Advanced

Settings

When the INFO key is pressed the display looks like
Figure 5-11:
Using the keys, move the pointer to indicate:

System Information, Hardware Settings, or

Advanced Settings

Press ENTER to make the selection.

1.

System Information

Manufacturer’s information is shown including the
manufacturer’s name, the name of the system (iQ 20kHz
Hand Probe) and the software identification.

Figure 5-11A shows an example of this information.

2.

Hardware Settings

Figure 5-12 gives a view of the Hardware Settings
screen.

Audible Alarms - With the pointer as shown in Figure
5-12, the operator can turn the audible alarms on or
off. Use the keys to move from ON to OFF.
The current selection is highlighted.

Latching Faults - Latching faults can be turned on
or off.
If Latching Faults is on, and a fault occurs, the alarm
sounds twice (beep, beep) indicating the generator
is latched and it will not weld again until the fault is
corrected.
Pressing ENTER clears the fault, and the next cycle
can begin.

If Latching Faults is off, and a fault occurs, the audible
alarm (beep, beep) will be heard, but welding can
continue.

Continued

iQ Information

> System Information

Hardware Settings

Advanced Settings

Figure 5-11 INFO Screen

Hardware Settings

>

Audible Alarm

ON

OFF

ON

OFF

Press Enter To Accept

Press Cancel To Abort

Latching Faults

Figure 5-12 Hardware Settings Screen

+ -

NOTE

Changes in Hardware Settings take
place after ENTER is pressed.

Figure 5-11A System Information Example Screen

Dukane Corporation
iQ 20kHz Hand Probe
FP v1.07 MB v2.56
FPGA v6

Advertising