Section 7 – Troubleshooting

Page  47

Dukane Manual Part No. 403-577-01

No Ultrasonic Output

Probe

Make sure that the hand probe cable is connected to the
generator connector (HAND PROBE) and secured to
the rear panel. Also, make sure the hand probe stack is
properly assembled.

System Power Output Level

Overload

When an overload occurs, it will automatically reset
when the next ultrasound activation signal begins. If the
condition persists:

Turn the generator OFF and:
1. Check the system. Change the hand probe to one that

is known to be good.

2. Turn the generator

ON, and see if the fault condition

has been corrected.

Overtemperature

When the system overheats, and the generator’s internal
temperature exceeds 85° C (185° F) an overtemperature
fault condition will trip.
When the system cools, the system automatically resets
the overtemperature fault.

Generator Fault Does Not Reset

When a system fault condition does not automatically
reset, the generator needs servicing.

System Power Diagnostic
Procedures

The only fault indications available with these probe
systems are the ones shown on the LCD display.

NOTE

The LCD screen displays a vari-
ety of pop-up status changes as
they occur. Check Tables 7-I and
7-II - Pop-up Fault Status Screens
- Pages 49- 50.

NOTE

When Latching Faults is enabled,
ENTER must be pressed to clear
a fault.