Flowserve 3200MD User Manual

Page 19

Advertising
background image

User Instructions - Digital Positioner 3200MD LGENIM0059-09 12/13

19

Blink

Code

Description

Recommendations

**YYGR SUPPLY PRESSURE HIGH WARNING indicates the positioner

has determined that the supply pressure is above the user set

warning limit.

Regulate the supply pressure at the positioner below the maximum limit

recommended for your actuator. Recalibrate pressure sensors. Check

the pressure sensor board connections. Replace pressure sensor board

if necessary. See DTM screen: Health Status/Actuator Health. This

indicator can be disabled.

**YYYG SUPPLY PRESSURE LOW WARNING (user set) indicates that the

supply pressure is below the user set warning limit. Low supply

pressure can cause poor valve response or positioner failure.

The minimum recommended supply pressure is 30 PSI (2.1 bar)

for proper operation. The unit will fail at less than approximately

17 PSI (1.2 bars). Low supply pressure indications can also be

caused by pneumatic leak.

Regulate the supply pressure at the positioner above 30 PSI (2.1

bar). Recalibrate pressure sensors. Ensure system air/gas supply is

adequate. Repair kinked supply tubing. Check the pressure sensor board

connections and replace pressure sensor board if necessary. Check for

pneumatic leaks in the actuator and actuator tubing. See DTM screen:

Health Status/Actuator Health. This indicator can be disabled.

**YYYY ACTUATION RATIO WARNING (user set) indicates a decreased

ability of the system to actuate the valve. It is based on the ratio

of available force to required force to actuate. It is affected by the

process load, friction, spring force, and available supply pressure.

Increase the supply pressure. Reduce the friction. Check the actuator

spring. Resize the actuator. Adjust user set limits. See DTM screen:

Health Status/Actuator Health. This indicator can be disabled.

**YRGG PILOT RELAY RESPONSE WARNING (user set) indicates that

the pilot relay is sticking or slow to respond. This affects the

responsiveness, increases the chance of limit cycling and

excessive air consumption. The pilot relay is part of the inner

loop and consists of the driver module assembly with piezo (I-P

relay) which is coupled to the spool valve. The value of this

indicator corresponds with inner loop lag. Delayed response can

be caused by a partially clogged piezo or debris, oil, corrosion, or

ice on the spool, or low supply pressure.

Check response of the valve. If OK, adjust Pilot Relay Response limits.

Check supply pressure. Check the spool for debris, oil, corrosion, ice

on the spool. Clean or replace the spool assembly. Replace the piezo

or driver module assembly. Maintain a clean, water-free air/gas supply.

See DTM screen: Health Status/Positioner Health. This indicator can be

disabled.

**YRGY FRICTION LOW WARNING (user set) indicates the friction has

passed below the user set limit.

Low friction is usually an indication of improperly loaded packing or seals

in the valve and actuator. See DTM screen: Health Status/Valve Health. This

indicator can be disabled.

**YRGR PNEUMATIC LEAK WARNING (user set) indicates that the positi-

oner has detected a leak in the actuation assembly. Leakage from

the actuator can cause decreased responsiveness and excessive

air/gas consumption. Low supply pressure can also trigger this

warning.

Repair pneumatic leaks at the tubing junctions and actuator seals. Ensure

proper supply pressure. See DTM screen: Health Status/Actuator Health.

This indicator can be disabled.

YRYG

FRICTION HIGH WARNING (user set) indicates the valve/actuator

friction has passed the user set limit. High friction can cause loop

oscillations, poor position control, jerky motion, or valve sticking.

It can be caused by build-up from the process on the stem, trim or

seat , by a failing bearing or guides in the valve and actuator,

galling of the trim or stem, excessively tightened packing, linkages,

or other valve/actuator mechanical issues.

Determine if the friction is signifi cantly interfering with the valve control. If

not, consider increasing the friction warning limit. Consider the following

to reduce friction: Stroke the valve to clear off build-up. Clear any external

mechanical obstruction, loosen the packing, clean the stem, repair or re-

place the actuator. Highly localized friction or very jerky travel can indicate

internal galling. Repair or replace internal valve components. See DTM

screen: Health Status/Valve Health. This indicator can be disabled.

YRRY

ELECTRONIC INABILITY TO FAIL SAFE WARNING indicates that

the piezo may be damaged. This may prevent the proper failure

position upon loss of signal/power. This condition may occur

briefl y on an air-to-close valve that is held for long periods of

time in the closed position, or and air-to-open valve held in the

open position.

If alarm persists for more than 30 minutes, the Piezo assembly is dama-

ged and should be replaced. This indicator can be disabled.

**YRRR PNEUMATIC INABILITY TO FAIL SAFE WARNING indicates that

upon loss of air supply, the valve may not move to the fail-safe

position. The spring alone is not adequate to overcome the

friction and process load in the system. The system is relying on

pneumatic force to actuate in the direction the spring is pushing.

The failsafe spring may have failed, or it was not sized properly

for the application. Friction or process load may have increased.

Check for high friction. Repair or replace actuator spring. Reduce pro-

cess load. This indicator can be disabled.

RGGY

FEEDBACK READING PROBLEM DURING CALIBRATION ALARM

indicates that during calibration, the range of motion of the positi-

on feedback arm was too small for optimum performance, or the

position sensor was out of range.

Check for loose linkages and/or adjust the feedback pin to a position

closer to the follower arm pivot to create a larger angle of rotation if the

feedback rotation is less than 15 degrees for the total valve travel and

recalibrate. Briefl y pushing the QUICK-CAL button acknowledges this

condition and the positioner will operate using the current short stroke

calibration if otherwise a good calibration. If the condition does not clear

then adjust the positioner mounting, linkage or feedback potentiometer

to move the position sensor back into range then restart the calibration.

This error may be cleared by briefl y pushing the QUICK-CAL button,

which will force the positioner to use the parameters from the last good

calibration.

Advertising