14 frequently asked questions – Flowserve 3200MD User Manual

Page 37

Advertising
background image

User Instructions - Digital Positioner 3200MD LGENIM0059-09 12/13

37

14 Frequently Asked Questions

Q: My DCS uses 24 VDC, can I run a Logix 3200MD?

A: A DCS output current card does run from 24 VDC but the card regulates

the actual current output. However, if the 24 VDC were applied directly across

the terminals, nothing would limit the current and the Logix 3200MD Field

Termination Board could be damaged. The current source voltage supply can

be anywhere between 10 VDC to 30 VDC as long as the current is limited in the

4-20 mA range.

Q: I accidentally placed a voltage supply across the Logix 3200MD. How do

I know if I damaged something?

A: The typical failure in an over-current situation is a short circuit. Your loop

current will be maintained but the Logix 3200MD control board will receive

no power. With power removed from the Logix 3200MD, use an ohmmeter to

measure across the terminals. If the reading is a short (close to zero ohms),

the Customer Interface Board must be replaced. Make sure the positive lead is

on the ‘+’ terminal and negative lead is on the ‘-’ terminal when measuring the

resistance.

Q: What is the input resistance of the Logix 3200MD?

A: The Logix 3200MD does not have a simple resistive input. When measuring

the voltage across the Logix 3200MD, it only varies slightly when the current

is changed from 4 mA to 20 mA (9.8 to 10.0 VDC nominal without HART com-

munications. Add 0.3 V with HART active). This is because the Logix 3200MD

is an active device. The resistance at a given current is commonly referred to

as the effective resistance.

Effective Resistance = (Terminal Voltage)/Current

Equation 3

For example:

at 20 mA: Effective resistance = 9.9 VDC/0.02 A = 495 Ω

The Logix 3200MD has a specifi cation of 495 Ω @ 20 mA.

!

NOTE: Your cannot measure across the terminals of an unpowered Logix

3200 and get the effective resistance.

Q: How do I know if I need a VHF HART fi lter?

A: If the current source is interfering with communication, it will affect both

ValveSight and the HART 375 handheld. If the positioner communicates to

ValveSight or the HART 375 handheld when using a current source (a 4-20 mA

current calibrator, for example) and not the DCS, it indicates a fi lter is neces-

sary with that current source. Some 4-20 mA calibrators that work without a

fi lter are listed below. If one of these is available, try to connect with ValveSight

or the HART 375 handheld again. If communications are established while us-

ing one of these sources but fails on the original source, a fi lter is needed.

Handheld 4-20 mA calibrators which do not require a fi lter:

• Altek Model 334

• Rochester Instrument Systems (RIS) CL-4002

• Unomat UPS-II

Q: I set the MPC at 5 percent. How will the positioner operate?

A: Assume that the present command signal is at 50 percent. If the command

signal is decreased, the positioner will follow the command until it reaches

5 percent. At 5 percent, the spool will be driven fully open or fully closed,

depending on the air action of the valve, in order to provide full actuator

saturation and tight shutoff. The positioner will maintain full saturation below

5 percent command signal. As the command increases, the positioner will

remain saturated until the command reaches 6 percent (there is a 1 percent

hysteresis value added by the positioner). At this point, the stem position will

follow the command signal. While in MPC, the Logix 3200MD LEDs will blink

GGGY.

Q: I have MPC set to 3 percent but the valve will not go below 10 percent.

A: Is a lower soft stop enabled? The lower soft stop must be less than the

lower PMC setpoint in order for the lower MPC to become active. When the

lower soft stop is reached, the positioner will blink a GYGY code. When MPC

is active, the positioner will blink GGGY.

Q: Will soft stops prevent the valve from going to its fail position?

A: No.

Q: What is the difference between models with Standard (3200MD),

Advanced (3210MD) and Pro (3220MD) diagnostics?

A: All models use the pressure sensors in the positioner control algorithm

to enhance valve stability. The model with Advanced diagnostics performs

calculations such as detection of loss of pressure, diagnostic signatures and

data logging. The model with Pro diagnostics adds full diagnostic capabili-

ties including continuous on-line friction, pneumatic leak and actuation ratio

analysis. Pro diagnostics support comprehensive health views in the ValveS-

ight DTM. Partial stroke tests and long term trends of many key parameters

are also available with Pro diagnostics.

Q: Can I upgrade from a Standard to Advanced or from Advanced to Pro

diagnostics?

A: Yes. Upgrading can be accomplished using ValveSight DTM software.

After upgrading, pressure and friction calibrations can be performed using

the ValveSight DTM or the HART 375 handheld.

Q: Can I add pressure sensors to my 3200IQ and upgrade it to a 3220MD

with Pro diagnostics?

A: This depends on the certifying agency. Some agencies do not allow fi eld

upgrades which change the physical confi guration. If the certifying agency

allows physical upgrades, the pressure sensor board assembly can be pur-

chased and installed. A Flowserve technician can update the positioner soft-

ware from IQ to the desired MD version (Advanced or Pro). If the certifying

agency does not allow physical upgrades, a Flowserve technician can update

the positioner software from the IQ to the desired MD version (Advanced or

Pro). If the certifying agency does not allow physical upgrades, a Flowserve

technician can update the positioner software to turn the 3200IQ into a

3200MD (with Standard diagnostics) allowing it to work with the ValveSight

DTM. See the Logix 3200IQ IOM for instruction on removing the plug plate

and installing the pressure sensors. For any questions regarding this upgrade

or the certifying agencies’ rules, contact the Flowserve Springville Quality

Department.

Q: Does the QUICK-CAL perform all the required calibrations for an Ad-

vanced or Pro Diagnostics level Logix positioner?

A: No, the advanced diagnostics feature of the ValveSight technology requires

the user to input the size of the actuator and the supply pressure into the

positioner. This cannot be done with a QUICK-CAL only. Perform a Stroke,

Actuator and Friction calibration through the ValveSight DTM or HART

handheld.

Advertising