Fluke Biomedical 4000M+ User Manual

Page 38

Advertising
background image

Victoreen 4000M+
Operators Manual

B-4

The following is a sample program:

5 REM THIS PROGRAM READS THE FIRST HUNDRED DATA
6 ‘ POINTS INTO THE ARRAYS A AND B.
10 PINT #1, W;
20 FOR I=1 TO 100 STEP 10
30 PRINT #1, I
40 FOR J=I TO I+10
50 IPUT #1,A(I), B(I)
60 NEXT J
70 NEXT I
80 PRINT #1, CHR$(27);

Fluoroscopic Measurement (U)

The 4000M+ may be used in the Fluoroscopic Mode while under control of a host computer. The
procedure is as follows:

1. Send the character U to the 4000M+ via RS-232.
3. The 4000M+ responds with a status code; a zero (0) indicates all is ok; any other code indicates a

hardware error.

4. Once per second, thereafter, the 4000M+ sends one or two numbers, followed either by R or K and

then a carriage return/line feed.

The numbers followed by R are exposure rate, in R/min. Those followed by K are kilo voltage
values. When there is little or no exposure, only the R values will be given.

5. This output will continue indefinitely. To stop it, repeatedly send the character X until the 4000M+

responds.

NOTE


The 4000M+ does not look at the RS-232 port while
it is accumulating real-time data. Therefore, send
the X repeatedly until 4000M+ responds.

6. The 4000M+ responds by repeatedly sending 3 <CR><LF> for a period of time, giving the computer

a chance to alternate between sending the X and looking for the response.

7. When the 3s are no longer being sent, the 4000M+ will have accumulated a final 100 ms of

waveform data. A final status code is sent. A code of 01 means all has gone well. A code of 50,
51, or 53 indicates low kVp, high kVp, or all data below radiation threshold, respectively.

8. The

commands

D

or W may now be used to review the data.

Set Detector Sensitivity (H or L)

Set the detector sensitivity as follows:

1. Send the character H to the 4000M+ to set the sensitivity to high.
2. Send the character L to the 4000M+ to set the sensitivity to low.


The 4000M+ responds by echoing back the character sent and an 01 to indicate Ready.