Section 2, Operation, 1 power – Fluke Biomedical 4000M+ User Manual

Page 9: 2 checking the unit

Advertising
background image

2

Operation

Power

2-1

Section 2

Operation


2.1 Power

WARNING

Avoid shorting the battery. Do not mutilate or put a
fire. Use only approved charging methods.

The Model 4000M+ can be powered by the internal rechargeable nickel cadmium batteries or by AC line
power. An AC adapter is supplied with each unit. The connector for the adapter is located on the rear
panel and labeled POWER (see Figure 2-2).

The batteries charge whenever the unit is plugged in to an AC source. The batteries will fully charge in
approximately 6 hours, with the unit turned off, or in 24 hours, with the unit turned on. The batteries are
protected from overcharging.

2.2 Checking the Unit

Use the following procedure to check the Model 4000M+ for proper operation in the Radiographic Mode:

1. Place the unit on the x-ray table with the black plastic disk facing up.
2. Turn the Power Switch to ON. The Model No., software revision level, and filter setting will be

displayed, followed by a FLUORO Ready or RADIO Ready indication, depending on the position of
the FLUORO/RADIO Switch.

3. Position the Mode Selection Switch to RADIO.
4. Position

the

EXP/ALL

Switch to ALL.

5. Position the Tube Target (Anode)/Filter Selection Switch to W/Al.
6. Position the Sensitivity Selection Switch to LOW.
7. Rotate the filter wheel to the 70-120 kVp setting. The RADIO Ready indication should appear on

the display.

NOTE


If there is a hardware failure, an ERROR message
may be displayed instead of the RADIO Ready
condition. Error message indications and
meanings are discussed in Section 3 – Theory of
Operation.

8. Position the detector as discussed in Section 3 – Theory of Operation.
9. Make an exposure at 80 kVp and 100 mA and 100 mS.
10. Ready the display.