Victoreen 4000M+
Operators Manual

1-2

1.2 Specifications

Measured Quantities

Measured during the first 300 ms of exposure:
- kVp average
- kVp effective
- kVp maximum
Accuracy: +2%
Range:
W/Al: 27 to 155 kVp, calibrated to a tungsten anode tube with
4.5 mm Al total filtration;

Kilovoltage

Mo/Mo: 21-50 kVp, calibrated to a molybdenum anode tube
with 0.03 mm Mo total filtration;

Accuracy: 1 kV Mo/Mo (22 to 35 kVp)
(Mammo generators w/30 µ Mo)
Measured during entire exposure; referenced to 90% rise/fall
kV time.
Accuracy: Within 2% or 2 ms, which ever is greater.

Time

Range: 1 ms to 10 sec.
Measured during entire exposure, kVp corrected.
Accuracy: +5%.

Exposure

Range: 10 mR to 10 R
Accuracy: +5%

Fluoroscopic Rate

Range: 0.5 to 200 R/min.

Detection Methods

kV

CsI/photodiode pair measures x-ray transmission through
differential copper attenuator.

Time

Computed from kV waveform stored in memory, against
quartz crystal time base.

Exposure

Parallel plate ionization chamber (36 cm3 volume).

Calibration Check

Reference to a calibrated voltage divider and a calibrated
exposure monitor during irradiation.