Fluke Biomedical ESA609 User Manual

Page 15

Advertising
background image

Electrical Safety Analyzer

Safety Information

3

 Warning

To prevent possible electrical shock, fire, or
personal injury, follow these guidelines:

•

Read all safety information before you use

the Product.

•

Use the Product only as specified, or the

protection supplied by the Product can be

compromised.

•

Do not connect to live output terminals.

The Product can supply voltages that can

cause death. Standby mode is not

sufficient to prevent electrical shock.

•

Do not apply more than the rated voltage,

between the terminals or between each

terminal and earth ground.

•

Limit operation to the specified

measurement category, voltage, or

amperage ratings.

•

Use Product-approved measurement

category (CAT), voltage, and amperage

rated accessories (probes, test leads, and

adapters) for all measurements.

•

Measure a known voltage first to make

sure that the Product operates correctly.

•

Use the correct terminals, function, and

range for measurements.

•

Do not touch voltages > 30 V ac rms, 42 V

ac peak, or 60 V dc.

•

Do not use the Product around explosive

gas, vapor, or in damp or wet

environments.

•

Do not use the Product if it is damaged.

•

Disable the Product if it is damaged, by

removing the power cord and fuses.

•

Do not use the Product if it operates

incorrectly.

•

Examine the case before you use the

Product. Look for cracks or missing

plastic. Carefully look at the insulation

around the terminals.

•

Do not use test leads if they are damaged.

Examine the test leads for damaged

insulation, exposed metal, or if the wear

indicator shows. Check test lead

continuity.

•

Use this Product indoors only.