Electrical Safety Analyzer

Performing Electrical Safety Tests

21

Performing a Direct-Equipment Leakage Test

Note

The Direct-Equipment Leakage test is available
when the EN62353 standard is selected.

The Direct-Equipment Leakage Current test measures the
leakage current between the exposed conductive surface
on the housing, to mains earth.

To perform a direct-equipment test, press .

The direct-equipment test is the default test and should
already be selected.

Figure 12 shows the electrical connections between the
Analyzer and the DUT during a Direct-Equipment Leakage
Current Test.

The following outlet conditions apply when performing this
test:

Normal Polarity, Closed Earth

Normal Polarity, Open Earth

Reversed Polarity, Closed Earth

Reversed Polarity, Open Earth