ESA609
Users Manual

16

Performing a Current Consumption Test

To measure the current consumed by the DUT, press .
The Analyzer displays the current flowing through the
mains connections of the test receptacle.

Performing Leakage Current Tests

The Analyzer measures leakage current for a number of
different DUT configurations.

The leakage tests that are available depend on which
standard is selected. See the “Selecting the Test
Standard” section earlier in this manual to change the
standard the Analyzer is using.

Table 5 lists tests that have different names based on
which standard is selected.

Table 5. Test Names Based on Selected Standard

IEC60601 AAMI/NFPA

99

Protective-Earth
Resistance

Ground-Wire Resistance

Earth-Leakage Current

Ground-Wire Leakage
Current

Touch- or Enclosure-
Leakage Current

Chassis-Leakage Current

Press to access the leakage current main test shown in
Figure 9.

fis217.bmp

Figure 9. Leakage Current Test

Note

The display shown in Figure 9 is a leakage current
test when AAMI is the selected standard.

All leakage currents are displayed in as RMS only. The
initial result is displayed in the appropriate parameter
based on the standard selected.