A low resistance reading is required to confirm a good
ground connection through the power cord. Refer to the
appropriate electrical safety standard for the specific limit
value to be followed.

Figure 8 shows the electrical connections between the
Analyzer and the DUT. Table 4 lists the abbreviations used
in the schematics and their descriptions.

Table 4. Schematic Abbreviations

Abbreviation Meaning

MD

Measuring Device (ESA609 Analyzer)

FE Functional

Earth

PE Protective

Earth

Mains

Mains Voltage Supply

L1 Live

Conductor

L2 Neutral

Conductor

DUT

Device Under Test

DUT_L1

Device Under Test Live conductor

DUT_L2

Device Under Test neutral conductor

DUT_PE

Device Under Test protective earth

REV POL

Reversed mains supply polarity

PE Open

Open protective earth

Test

Voltage