Electrical Safety Analyzer

Making Point-To-Point Measurements

23

Making Point-To-Point Measurements

The Analyzer can make resistance and low current
measurements through its Point-to-Point function. To
access the Point-to-Point menu press . Use the
softkeys F1 through F3 to select the measurement
function.

Measuring Resistance

To make a resistance measurement:

1. Press the softkey labeled Resistance from the

Point-To-Point menu.

2. Insert test leads in the red (

Ω

/A) and black jacks.

3. Null lead resistance by shorting the leads together and

pressing the softkey labeled Zero.

4. Place the probes across the unknown resistance and

read the measurement in the Analyzer’s display.

The Analyzer will measure resistances up to 20.0

Ω

.

Figure 13 is an example of the Resistance Test.

fis118.bmp

Figure 13. Point-to-Point Resistance Test

Measuring Leakage Current

The Analyzer can make true rms measurements up to
2 mA. To make a current measurement:

1. Press the softkey labeled Leakage from the

Point-To-Point menu.

2. Insert test leads in the red (

Ω

/A) and black jacks.

3. Place the leads on the two points the unknown current

may flow and read the measurement in the Analyzer’s
display.