Use only the mains power cord and

connector approved for the voltage and

plug configuration in your country and

rated for the Product.

Make sure the ground conductor in the

mains power cord is connected to a

protective earth ground. Disruption of the

protective earth could put voltage on the

chassis that could cause death.

Replace the mains power cord if the

insulation is damaged or if the insulation

shows signs of wear.

Connect the common test lead before the

live test lead and remove the live test lead

before the common test lead.

Keep fingers behind the finger guards on

the probes.

Do not use test leads if they are damaged.

Examine the test leads for damaged

insulation and measure a known voltage.

Do not use a current measurement as an

indication that a circuit is safe to touch. A

voltage measurement is necessary to know

if a circuit is hazardous.

Unpacking the Analyzer

Carefully unpack all items from the box and check that you
have the following items:

ESA609

Safety

Sheet

Users Manual CD

Carrying

case

Power

cord

ESA USA Accessory Kit (USA, Australia, and Israel

only)

ESA EUR Accessory Kit

Null Post Adapter