ESA620
Users Manual

16

Setting Polarity Switching Delay

When switching the polarity of the Analyzer’s test
receptacle, a delay can be set to control the actual switch
time. To set the polarity delay:

1. Press the softkey labeled Instrument from the setup

menu to reveal the instrument setup selections.

2. Press the softkey labeled Polarity Delay to open the

scroll box above the softkey label.

3. Press

 or  to adjust the delay from 0 to 5

seconds in 1 second steps.

4. Press the softkey Back then the softkey labeled Exit

to exit the setup function.

Setting the Display Contrast

There are two methods for setting the display contrast.
From the “Select a Test….” menu or through the setup
menu.

Whenever the Analyzer displays its start-up menu (Select
a test…), pressing  or  will increase or decrease the
display’s contrast respectively. Press the softkey labeled
Done to exit contrast setup.

Another way to adjust the contrast is through the
Analyzer’s setup menu.

1. Press the softkey labeled Display from the setup

menu.

2. Press the softkey labeled Contrast.

3. Press

 or  to increase or decrease the display’s

contrast respectively.

4. Press the softkey labeled Done to exit contrast

setup.

Setting up the Beeper

In addition to being enabled or disabled, the Analyzer’s
beeper volume can be set as well. To setup the beeper:

1. Press the softkey labeled Sound from the setup

menu.

2. Press the softkey labeled Beeper to switch the

beeper on and off.

3. Press the softkey labeled Volume to open a scroll

box above the softkey label.

4. Press

 or  to increase or decrease the volume

respectively.

5. Press the softkey labeled Done to go back to the

setup menu.