Performing electrical safety tests, Setting the test standard – Fluke Biomedical ESA620 User Manual

Page 31

Advertising
background image

Electrical Safety Analyzer

Performing Electrical Safety Tests

17

Performing Electrical Safety Tests

The Analyzer is designed to perform a number of different
electrical and performance tests on biomedical
equipment. The following sections describe the various
tests and how to perform them using the Analyzer.

Setting the Test Standard

The Analyzer is designed to perform electrical safety
testing based on a number of different safety standards.
The IEC 60601 is the Analyzer’s default standard. To
select another standard:

1. Press

.

2. Press

 or  to scroll through the standard

selections.

3. When the desired standard is displayed, press the

softkey labeled Select.

To exit the standard selection menu without changing the
standards selection, press the softkey labeled Exit.

Some electrical tests may not be applicable for a specific
standard. In these cases, the Analyzer’s menu will not
display the excluded test as a selection.

Performing an Accessible Voltage Test (IEC
61010 only)

Note

The Accessible Voltage test selection will only
appear in the Analyzer’s menu when the
standard is set to IEC61010.

The Accessible Voltage test measures the voltage that
may exist between the DUT’s enclosure and protective
earth. To access the Accessible Voltage test, press .

1. Connect the DUT’s power cord to the Analyzer’s test

receptacle.

2. Connect a test lead from the Analyzer’s 2-WIRE

V/

Ω

/A jack and an exposed metal part on the DUT’s

enclosure. Any measured voltage is shown in the
Analyzer’s display.

The following outlet conditions apply when performing this
test:

Normal

polarity

Normal polarity, earth open

Normal polarity, neutral open

Reversed

polarity

Reversed polarity, earth open

Reversed polarity, neutral open