Electrical Safety Analyzer

Detailed Specifications

59

Additional error caused by series inductance

Resistance

Series Inductance

0 µH

100 µH

200 µH

400 µH

0.000

Ω

0.000

Ω

0.030

Ω

0.040

Ω

0.050

Ω

0.020

Ω

0.000

Ω

0.025

Ω

0.030

Ω

0.040

Ω

0.040

Ω

0.000

Ω

0.020

Ω

0.025

Ω

0.030

Ω

0.060

Ω

0.000

Ω

0.015

Ω

0.020

Ω

0.025

Ω

0.080

Ω

0.000

Ω

0.010

Ω

0.015

Ω

0.020

Ω

0.100

Ω

0.000

Ω

0.010

Ω

0.010

Ω

0.015

Ω

>0.100

Ω

0.000

Ω

0.010

Ω

0.010

Ω

0.010

Ω

Equipment Current

Range .................................................................. 0 – 20 A ac rms

Accuracy.............................................................. 5 % of reading

±

(2 counts or 0.2A, whichever is greater)

Duty cycle ............................................................ 15 A to 20 A, 5 min. on/5 min. off

10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous

Leakage Current

Modes* ................................................................ AC+DC (True-rms)

AC only
DC only
* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP
that are available in True RMS (shown as AC+DC)

Patient Load Selection ........................................ AAMI ES1-1993 Fig. 1

IEC 60601: Fig 15
IEC 61010: Fig A-1