Performing a direct applied part leakage test – Fluke Biomedical ESA620 User Manual

Page 59

Advertising
background image

Electrical Safety Analyzer

Performing Electrical Safety Tests

45

Performing a Direct Applied Part Leakage Test

Note

The Direct Applied Part Leakage test is available
when the EN62353 & VDE 751 or the None
standard is selected.

The Direct Applied Part Leakage Current test measures
the leakage current between all applied parts of a single
function and the exposed conductive surface on the
housing, to mains earth. For equipment with multiple
applied parts, each group of a single function should be
tested each in turn with all other floating during the test.
This test should only be done for equipment with F-Type
applied parts.

For Type B applied part, see direct equipment leakage
schematic in Figure 27.

To perform a direct applied part leakage test:

1. Press

.

2. Press the soft key labeled More.

3. Select the desired applied part groupings using

 and .

4. Press the soft key labeled Select. The Direct A.P.

test should already be selected.

5. Press

 or  to select the applied part test

configuration.

6. Press

 to apply the test voltage and read the

current in the display.

7. Press

 or  to advance to the next group of applied

parts, if applicable.

Figure 28 shows the electrical connections between the
Analyzer and the DUT during a Direct Applied Part
Leakage Current Test.

The following outlet conditions apply when performing this
test:

Normal

Polarity

Reversed

Polarity