Detailed specifications – Fluke Biomedical ESA620 User Manual

Page 72

Advertising
background image

ESA620
Users Manual

58

Detailed Specifications

Voltage

Mains voltage

Ranges ............................................................ 0.0 to 300 V ac rms

Accuracy .........................................................

±

(2 % of reading + 1.0 V ac)

Accessible Voltage and Point to Point Voltage

Range ............................................................. 0.0 to 300 V ac rms

Accuracy .........................................................

±

(2 % of reading + 2 LSD)

Earth Resistance

Modes ................................................................. Two terminal and four terminal

Test Current ........................................................ >200 mA ac into 500 m

Ω

with open circuit voltage

24 V

25 A short circuit

±

10 % (with open circuit voltage 6 Vac at nominal mains)

Range ................................................................. 0.0 to 2.0

Ω

Accuracy

Two Terminal Mode

Test current >200 mA ac into 500 m

Ω

........

±

(2 % of reading + 0.015

Ω

) for 0.0 to 2.0

Ω

Test current 1-16 A ac ................................

±

(2 % of reading + 0.015

Ω

) for 0.0 to 0.2

Ω

±

(5 % of reading + 0.015

Ω

) for 0.2 to 2.0

Ω

Four Terminal Mode

Test current >200 mA ac into 500 m

Ω

........

±

(2 % of reading + 0.005

Ω

) for 0.0 to 2.0

Ω

Test current 1-16 A ac ................................

±

(2 % of reading + 0.005

Ω

) for 0.0 to 0.2

Ω

±

(5 % of reading + 0.005

Ω

) for 0.2 to 2.0

Ω