Electrical Safety Analyzer

Detailed Specifications

61

Insulation resistance

Ranges ................................................................ 0.5 to 20 M

Ω

20 to 100 M

Ω

Accuracy

20 M

Ω

Range ..................................................

±

(2 % of reading + 2 counts)

100 M

Ω

Range ................................................

±

(7.5 % of reading + 2 counts)

Source test voltage ............................................. 500 V dc (+20 %, -0 %) 1.5 mA short-circuit current or 250 V dc selectable

Maximum load capacitance ................................. 1

μ

F

ECG Performance Waveforms

Accuracy..............................................................

±

2 %

±

5 % for amplitude of 2 Hz square wave only, fixed @ 1 mV Lead II configuration

Waveforms

ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM

Ventricular Fibrillation

Square wave (50 % duty cycle) ....................... 0.125 and 2

Sine wave ........................................................ 10, 40, 50, 60, and 100 Hz

Triangle wave .................................................. 2 Hz

Pulse (63 ms pulse width ................................ 30 and 60