Vital Signs Simulator

Cardiac Functions

13

Pacemaker Signals

The Product can simulate heart signals with pacemaker
control signals. To set the waveform, amplitude, and width
of the pacemaker signal:

1. In

the

Home screen, push the navigation buttons to

highlight

PACE

.

2. Push

 to show the Pacemaker Waves screen in

the display.

3. Use the parameter selection procedure described in

the Operation section to set the parameters of the
pacemaker signal.

See the detailed specifications for the range of parameter
values.

Arrhythmia Function

The Product can simulate heart arrhythmias. To start an
ECG arrhythmia simulation:

1. In

the

Home screen, push the navigation buttons to

highlight

ARRY

.

2. Push

 to show the Arrhythmias screen in the

display. See Figure 10.

hal009.bmp

Figure 10. Arrhythmia Screen

The simulated arrhythmias are grouped into four
categories: Supraventricular, Premature, Ventricular,
and Conduction Defect. See the specifications to learn
the arrhythmias in each group.

3. Push

or  to move the highlight to an arrhythmia

group icon.

4. Push

.

5. Use the parameter selection procedure described in

the Operation section to set the arrhythmia.

When the type parameter has been set to an arrhythmia,
the group icon of the group the arrhythmia belongs to has
a thicker border around it.