General specifications, Detailed specifications, Ecg waveform – Fluke Biomedical ProSim 2 User Manual

Page 55

Advertising
background image

Vital Signs Simulator

General Specifications

43

General Specifications

Power ......................................................................... Two 9-V alkaline batteries (IEC 6LR61, NEDA 1604A). Optional battery eliminator: 15 Vdc, 1.5 mA
Battery Life ................................................................ 8 hours minimum
Display ....................................................................... LCD Greyscale Display
Size ............................................................................. 14.0 cm x 20.6 cm x 4.5 cm (5.5 in x 8.2 in x 1.8 in)
Weight ........................................................................ 0.47 kg (1 lb 4 oz)
Temperature

Storage ................................................................ -25

°

C to +50

°

C (-13

°

F to +122

°

F)

Operation ............................................................. 10

°

C to 40

°

C (50

°

F to 104

°

F)

Humidity .................................................................... 10 % to 80 % non-condensing
Altitude ...................................................................... 2000 m (6,562 ft)
Safety ......................................................................... IEC 61010-1, Pollution degree 2
Electromagnetic Environment ................................. IEC 61326-1, Portable
EMC ............................................................................ Applies to use in Korea only. Class A equipment (Industrial Broadcasting & Communication Equipment)

[1]

[1] This Product meets requirements for industrial (Class A) electromagnetic wave equipment and the
seller or user should take notice of it. This equipment is intended for use in business environments and
is not to be used in homes.

Detailed Specifications

ECG Waveform

ECG Reference.......................................................... The ECG amplitudes specified are for Lead II, from the baseline to the peak of the R wave. All other

leads are proportional in percentage per:

Lead I .................................................................. 70 %
Lead II ................................................................. 100 %
Lead III ................................................................ 30 %
Lead V1 ............................................................... 24 %
Lead V2 ............................................................... 48 %
Lead V3 ............................................................... 100 %
Lead V4 ............................................................... 120 %
Lead V5 ............................................................... 112 %