background image

Vital Signs Simulator

Remote Operation

27

When connected to a PC with a Windows operating
system, the Product will communicate through a PC COM
port. Make sure the COM port parameters are set to:

9600

Baud

No

Parity

8 data bits

1 stop bit

Hardware handshake set to off

Remote Commands

A remote command is made up of alphanumeric
characters. The first character of a command must be
alphabetic. The alphabetic characters can be upper or
lower case.

Special

characters

are:

Carriage return (CR)

Line feed (LF)

Space

(SP)

Backspace

(BS)

Escape

(ESC)

The Product will do a command when it receives a
carriage return and/or line feed. Alphabetic characters are
not case sensitive. When you type in a command, the
backspace deletes the last recorded character and the
escape key discards the complete command. When a
command is complete, the Product sends a response that
ends with a carriage return and line feed to the PC. Unless
other data is sent back from the Product, the response is
“OK” if the command is accepted by the Product. When a

command is not accepted by the Product, an error code
shown in Table 8 is sent to the PC.

Table 8. Error Codes

Error Code

Description

ERR=00

No commands allowed at this time

ERR=01 Unknown

command

ERR=02 Illegal

command

ERR=03 Illegal

parameter

ERR=04 Data

corrupted

ERR=05 Unknown

error

ERR=06

Option not installed

ERR=07 Incorrect

password

While the Product is operated from the front panel (local
mode) the remote interface will not respond to a command
until the command REMOTE is sent to the Product through
the USB port.