Fluke Biomedical ProSim 2 User Manual

Page 34

Advertising
background image

ProSim™ 2/3
Users Manual

22

To set a periodic FHR with IUP:

1. If the Fetal Maternal screen shows in the display, go to

step 3. If not, go to the Home screen of the Product
and push the navigation buttons to highlight

FE/MA

.

2. Push

 to show the Fetal Maternal screen.

3. Use the parameter selection procedure to set the FHR,

IUP, and Period parameters

4. Push the navigation buttons to highlight

START

.

5. Push

 to start the test. If the Period parameter is

set to Manual, the simulation stops automatically after
the IUP wave stops. Each time you push 
another IUP wave starts. If not set to Manual, the IUP
wave repeats at the frequency set in the Period
parameter until the simulation is stopped.

To stop the simulation, highlight

STOP

and push .