Fluke Biomedical ProSim 2 User Manual

Page 31

Advertising
background image

Vital Signs Simulator

Cardiac Functions

19

3. Set up the patient monitor to:

Catheter size: 7 F

Injectate volume: 10 cc

Injectate temperature: 0

°

C or 24

°

C

Computational constant: 0.542 for 0

°

C injectate

or 0.595 for 24

°

C injectate

4. In

the

Home screen of the Product, push the

navigation buttons to highlight

CO

.

5. Push

 to show the cardiac output screen.

6. Use the parameter selection procedure to set the

cardiac output parameters for the test. See the
detailed specifications to learn the range of each
parameter.

7. Push the navigation buttons to highlight

START

.

8. Push

 to start the test. The simulation stops

automatically.

To stop the simulation, highlight

STOP

and push .

How to Simulate Injectate Failure and Left-to-
Right Shunt Fault

The Cardiac Output function can simulate an injectate
failure or left-to-right shunt fault. To set either of these two
failures:

1. Push

or  to highlight the Wave value.

2. Push

or  until FAULTY INJ or L to R SHUNT

shows in the display.

3. Push the navigation buttons to highlight

START

.

4. Push

 to start the test.

How to Simulate Output from a Calibrated Pulse
Signal

The Product sources a waveform that simulates an
injectate temperature of 0

°

C or 24

°

C with a step of 1.5

°

C

for 1 second as a test for a cardiac-output monitor. To
output a calibration pulse:

1. In the Cardiac Output screen, push  or  to

highlight the Wave value.

2. Push

or  until CAL PULSE shows in the display.

3. Push

or  to highlight

START

.

4. Push

 to start the test.