ProSim™ 2/3
Users Manual

18

Blood temperature (BT) connector

Injectate temperature (IT)
thermistor connector

Injectate temperature (IT)
10-turn potentiometer

CI-3 Module/Cable

2

1

4

3

Pin 1

Pin 3

Crimp tube

Thermistor

end of cable

Assembly of the

General Purpose Connector

Viewed from

the top

of the C1-3

PIN

1

FUNCTION

INJECTATE THERMISTOR CONNECTION A

Injectate temperature (IT) thermistor connector

INJECTATE THERMISTOR CONNECTION B

2

NO CONNECTION

4

NO CONNECTION

3

Solder wires to
pins 1 and 3

hal010.eps

Figure 11. Cardiac Output Injectate CI-3 Adapter

For cardiac output simulation, use the supplied CI-3
adapter to connect the Product to the Device Under Test
(DUT). (see Figure 12). If necessary, use the general-
purpose connector.

To do a cardiac output test:

1. Connect the patient monitor to the cardiac output

adapter.

2. Connect the adapter to the Product (Figure 12).