Temperature, Cardiac output (prosim 3 only), Fetal / maternal-ecg (prosim 3 only) – Fluke Biomedical ProSim 2 User Manual

Page 59

Advertising
background image

Vital Signs Simulator

Detailed Specifications

47

Temperature

Temperature .............................................................. 0

°

C (32

°

F), 24

°

C (75.2

°

F), 37

°

C (98.6

°

F), and 40

°

C (104

°

F)

Accuracy ....................................................................

±

0.1

°

C

Compatibility ............................................................. Yellow Springs, Inc. (YSI) Series 400 and 700
Output Connector ..................................................... DIN 4-pin
Power-On Default...................................................... 0

°

C (42

°

F)

Cardiac Output (ProSim 3 Only)

Catheter Type ............................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient ............................................. 0.542 (0

°

C injectate), 0.595 (24

°

C injectate)

Blood Temperature ................................................... 37

°

C (98.6

°

F)

±

2 %

Injectate Volume ....................................................... 10 cc
Injectate Temperature .............................................. 0

°

C or 24

°

C

±

2 % value

Cardiac Output .......................................................... 2.5, 5, 10 liters per minute

±

5 %

Faulty-Injectate Curve .............................................. Waveform for simulation available
Left-to-Right-Shunt Curve ........................................ Waveform for simulation available
Calibrated Pulse........................................................ 1.5

°

for 1 second (37

°

to 35.5

°

)

Output Connector ..................................................... DIN 7-Pin
Power-On Default ..................................................... 2.5 liters per minute, 0

°

C injectate

Fetal / Maternal-ECG (ProSim 3 Only)

Fetal Heart Rate (Fixed)............................................ 60, 90, 120, 140, 150, 210, and 240 BPM
Fetal Heart Rate (IUP): .............................................. 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ........................... Early deceleration, late deceleration, and uniform acceleration
Wave Duration ........................................................... 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0 mmHg
IUP Period .................................................................. 2, 3, or 5 minutes, and manual
Power-On Default...................................................... FHR 120 BPM, early deceleration, manual