Multiparameter Simulator

Specifications

7

Blood Pressure

Input/Output Impedance .......................................350 Ohms
Exciter Input Limit
.................................................± 10 V
Exciter Input Frequency Range ...........................DC to 4000 Hz
Transducer Sensitivity..........................................5 or 40 µV/V/mmHg
Level Accuracy ......................................................± 1% ± 1 mmHg
Static Levels

BP 1.....................................................................-10, 0, 80, 160, 240, 320, and 400 mmHg
BP 2.....................................................................-10, 0, 50, 100, 150, 200, and 240 mmHg

Channel Selections

Arterial 120/80 .....................................................Channels 1 and 2
Radial Artery 120/80............................................Channels 1 and 2
Left Ventricle 120/00 ...........................................Channels 1 and 2
Right Ventricle 25/00 ...........................................Channels 1 and 2
Central Venous 15/10..........................................Channel 2
Pulmonary Artery 25/10.......................................Channel 2
Pulmonary Wedge 10/2.......................................Channel 2
Left Atrium 14/4 ...................................................Channel 2
Automatic Swan/Ganz......................................... Every 20 seconds
Manual Swan/Ganz ............................................ Changes every time entry is selected

Synchronized with all normal sinus rates

Physiologically tracks all arrhythmia selections