Fluke Biomedical PS420 User Manual

Page 39

Advertising
background image

Multiparameter Simulator

Simulating Functions

29

Code Display

Selects:

107

3L/M @ 2C

Cardiac output wave of 3
L/min for 2 degrees.
Computational constant is
0.561.

108

5L/M @ 2C

Cardiac output wave of 5
L/min for 2 degrees.
Computational constant is
0.561.

109

7L/M @ 2C

Cardiac output wave of 7
L/min for 2 degrees.
Computational constant is
0.561.

110

3L/M @ 20C Cardiac output wave of 3

L/min for 20 degrees.
Computational constant is
0.608.

111

5L/M @ 20C Cardiac output wave of 5

L/min for 20 degrees.
Computational constant is
0.608.

Code

Display Selects:

112

7L/M @ 20C

Cardiac output wave of 3
L/min for 20 degrees.
Computational constant is
0.608.

113

FLT INJ 2

Faulty injection cardiac output
wave for 2 degrees.
Computational constant is
0.561.

114

L- T SHT 2

Left to right shunt cardiac
output wave for 2 degrees.
Computational constant is
0.561.

115

FLT INJ 20

Faulty injection cardiac output
wave for 20 degrees.
Computational constant is
0.608.

116

L- R SHT 20

Left to right shunt cardiac
output wave for 20 degrees.
Computational constant is
0.608.