Pacemaker pulse amplitudes, lead ii, Pacemaker pulse width – Fluke Biomedical PS420 User Manual

Page 37

Advertising
background image

Multiparameter Simulator

Simulating Functions

27

Pacemaker Pulse Amplitudes, Lead II

When you select any pacemaker waveform, you can
adjust and set pulse amplitudes in Lead II. These can be
set through direct code entry, as below, or selected and
adjusted through first selecting 64 = PACE AMP. After
selecting 64, press ENTER, and then use the CHANGE
keys

to scroll to the available presets.

Code

Display

Sets pacemaker pulse

amplitude (Lead II) of:

224

PACE 2 mV

2 mV

225

PACE 4 mV

4 mV

226

PACE 6 mV

6 mV

227

PACE 8 mV

8 mV

228

PACE 10 mV

10 mV

Pacemaker Pulse Width

When you select any pacemaker waveform, you can
adjust and set pulse width. These can be set through
direct code entry, as below, or selected and adjusted
through first selecting 65 = PACE WIDTH. After selecting
65 press ENTER, and then use the CHANGE keys

to scroll to the available presets.

Code

Display

Sets pacemaker pulse

width of:

229

PACE 0.1mS

0.1 ms

230

PACE 0.5mS

0.5 ms

231

PACE 1 mS

1.0 ms

232

PACE 1.5mS

1.5 ms

233

PACE 2 mS

2.0 ms