Channel bp-1, Channel bp-2 – Fluke Biomedical PS420 User Manual

Page 30

Advertising
background image

PS420
Users Manual

20

Code Display

Sets

207

BP STAT L0 BP channel to static level 0.P1= -

10 P2= -10 P3= -5 P4= -5

208

BP STAT L1 BP channel to static level 0.P1=0

P2=0 P3=0 P4=0

209

BP STAT L2 BP channel to static level

0.P1=80 P2=50 P3=20 P4=20

210

BP STAT L3 BP channel to static level 0.

P1=160 P2=100 P3=40 P4=40

211

BP STAT L4 BP channel to static level 0.

P1=240 P2=150 P3=60 P4=60

212

BP STAT L5 BP channel to static level 0.

P1=320 P2=200 P3=80 P4=80

213

BP STAT L6 BP channel to static level 0.

P1=400 P2=240 P3=100
P4=100

Channel BP-1

Code

Display Selects:

78

P1 ARTERIA

Arterial waveform, 120/80.

79

P1 RADIAL

Radial waveform, 120/80.

80

P1 LVNT

Left ventricle waveform, 120/00.

81

P1 RVNT

Right ventricle waveform, 25/00

Channel BP-2

Code

Display Selects:

86

P2 ARTERIA

Arterial waveform, 120/80.

87

P2 RADIAL

Radial waveform, 120/80.

88

P2 LVENT

Left ventricle waveform, 120/00.

89

P2 RVNT

Right ventricle waveform, 25/00.

90

P2 PULAR

Pulmonary arterial waveform,
25/10.

91

P2 PULWDG

Pulmonary wedge waveform,
10/2.

92

P2 L ATRIU

Left atrium waveform, 14/4.

93

P2 CVP

Central venous waveform (right
atrium), 15/10.

Advertising