Ecg performance testing – Fluke Biomedical PS420 User Manual

Page 40

Advertising
background image

PS420
Users Manual

30

Code Display

Selects:

117

CAL WAVES Calibrated cardiac output

waves.

118 NONCAL

WVE

Uncalibrated cardiac output
waves. Gives 4 different
values per 3, 5, and 7 L/min
waves.

119

CAL 1 SEC

Calibration pulse of 1 degree
for 1 second.

120

CAL 4 SEC

Calibration pulse of a delta of
402 Ohms for 4 seconds.

121

SET BT 37C 37 °C (98.6 °F)

122

SET BT 36C 36 °C (95.9 °F)

When the monitor under test indicates that it is ready to
perform a Cardiac Output calculation, press ENTER on
the Simulator to generate a test. The 4-second calibration
pulse provides a delta change of 402 Ohms. This
provides a standard for testing cardiac output units.

If you select any cardiac output waveform key a second
time while a waveform is proceeding, the selected
waveform will stop. The blood temperature will return to
37 °C (98.6 °F.)

ECG Performance Testing

Square Wave

Code

Display Selects:

128

2 Hz SQR

2.0 Hz square waveform.

129

.125 Hz SQ

0.125 Hz square waveform.

Triangle Wave

Code

Display Selects:

130

2 Hz TRIAN

2.0 Hz triangle waveform.

131

2.5 Hz TRI

2.5 Hz triangle waveform

Pulse Wave

Code

Display Selects:

132

PULSE 30

Pulse of 30 BPM, width of 60 ms.

133

PULSE 60

Pulse of 60 BPM, width of 60 ms.