Formax FD 150 User Manual
Page 43

Operating Manual
1/97
39
WARNING
A BLOWN FUSE IN THE NEUTRAL LEG COULD
MEAN THAT INTERIOR PARTS OF THE
FD 150
REMAIN AT A HAZARDOUS VOLTAGE. ALWAYS
UNPLUG THE POWER CORD BEFORE SERVICING
THE
FD 150.
Figure 7.3 -- Replacing A Fuse
1. Unplug the POWER CORD.
2. Use a small screwdriver or similar tool to push up on
and release the FUSE DRAWER LOCKING TAB.
3. Pull the FUSE DRAWER out of the POWER ENTRY
CASING.
4. Inspect the FUSES; look for blackened glass, melted
wire or a disconnected wire between the ends of the
tube. If you find any of these problems in either FUSE,
that FUSE is blown and needs to be replaced.
5. Pull the blown FUSE from its slot.
6. Place a new FUSE into the same slot. Both fuses must
be in place for the FD 150 to operate properly.
7. Install the FUSE DRAWER in the POWER ENTRY
CASING.