Inova Labs XYC100B-P4L User Manual

Page 16

Advertising
background image

14

OPERATING PROCEDURE

1. Position the device so that the Fan Inlet Vent and Fan Outlet Vent are

not obstructed and in a well-ventilated space.

2. Briefly press the Power Button to

turn the device on. The Display
Screen and green Power Indicator
will illuminate. This indicates that
the device is powered on, ready for
use, and the settings may now be
changed.

NOTE: After powering on the device, allow 3 minutes for the system

to reach stated performance. If used routinely, the Pulse Setting last
used will appear on the display after powering on.

3. Use the Plus/Minus Buttons to select

the Pulse Setting prescribed by your
doctor (1, 2, 3 or 4). To change a
setting at any time, use the Plus and
Minus buttons.

4. Place the nasal cannula onto the Cannula Nozzle Fitting on the

device and the cannula prongs into your nasal passages and
breathe normally. When the device senses inhalation, oxygen is
supplied through the cannula. The duration and size of the pulse
is dependent upon the Pulse Setting selected.

NOTE: The device utilizes PULSE-WAVE Delivery which delivers

oxygen congruent with your inhalation rate. Because this type of
delivery is very gentle, you may not feel the pulse of oxygen
.
To verify that oxygen is being delivered, reference the symbol
next to your selected Pulse Setting on the display. This symbol will
flash every time a breath is detected.

5. Check to make sure the batteries are fully

charged before venturing out with the
device.

6. To turn the device off, press and hold the

Power Button until you hear an audible tone.

REPRESSURIZATION TECHNOLOGY

If the unit has not been used for a period of one or more days, it will
automatically re-pressurize itself. During this time, you may hear the
compressors running for a few seconds. There is no action required
on your part, this feature is intended to prolong the life of the device.