NORMAL OPERATION INDICATORS

During normal operation, the Control Panel and Display Screen
should appear as described below:

OPERATION

DESCRIPTION

Active or Rest Mode

Activitated

The word “Active Mode” or “Rest Mode” will appear next to your selected
pulse setting. If “Active Mode” is displayed it means your breath rate is typical
for an active user. If “Rest Mode” is displayed it means your breath rate is
slower than your normal active breath rate.

Alarm Muted

The audible alarm signal will be muted and the Mute symbol will
display on screen. The red alarm LED will remain illuminated on the
Control Panel and the alarm title will remain on the display.

Breath Detected

The PULSE-WAVE Breath Detection Symbol will flash as you breathe. This
is confirmation the device is delivering a dose of oxygen with each inhaled
breath.

If you stop breathing through the nasal cannula (mouth breathing), the unit
will begin delivering doses of oxygen automatically without the PULSE-
WAVE symbol. After 75 seconds, the unit will provide a “NO BREATH
DETECTED” alarm and will continue to automatically provide doses of
oxygen until normal breathing resumes. When the unit detects normal
breathing, the “NO BREATH DETECTED” alarm will cease and the PULSE-
WAVE symbol will indicate detection of your breath.

Charging Status

The following messages will appear on the display depending on the battery
being used or charged.
“Charging Internal” means the Internal Battery is charging.
“Charging External” means the External Battery is charging.
“Running Internal” means the Internal Battery is in use.
“Running External” means the External Battery is in use.

NOTE: By design, the device will preserve the Internal Battery charge as
long as possible. This means that the External Battery will deplete before the
Internal Battery and the Internal Battery will charge before the External Battery
when connected.

Connected to Activox DUO2 When the device is connected to the Activox DUO2 stationary base, the Display

Screen will show the message “Connected to Activox DUO2”.

Power Interruption

The device will emit a two second audible tone if the AC or DC Power is
interrupted or intentionally unplugged. The unit will automatically switch
to battery operation. If the power was not intentionally removed, check all
connections between the unit and the power source to ensure all are secure
and firmly plugged in. If the power indicator icon does not show on the
display, switch to another power source (AC or DC) or source of oxygen and
contact your equipment provider.

Powered On

The Power Indicator is illuminated green and the Display Screen is on.

Setting Selected

“1”, “2”, “3” or “4” will appear under “Setting” on the Display Screen.