18

ROUTINE CLEANING AND MAINTENANCE

The device was designed to minimize the amount of routine
maintenance that is required. There are no end-user repairable parts.
Aside from the preventative maintenance outlined below, all other
maintenance must be performed by qualified service personnel.

Fan Inlet Vent Cleaning: There is
one inlet vent located on the right
side (facing the unit) of the device.
The carrying case for the unit has
mesh material corresponding to the
location of the Fan Inlet Vent. You
should visually check to ensure that
there is no buildup of lint, hair or
other materials that could obstruct
the flow of air into the unit’s fan inlet
vent. At least once a month inspect
the grille of the intake vent and if it
appears dirty, use a dry cotton swab (Q-Tip or similar) to clean it.

Device and Carrying Case Cleaning: If it is necessary to clean the
device and/or carrying case, use only warm water and a mild liquid dish
detergent. Dampen a cloth in the solution and carefully wipe the
outside surfaces of the device and/or carrying case. Allow sufficient
time to dry before placing the device back in the case.

CAUTION: DO NOT use alcohol, solvents, polishes or any oily
substances as they may be flammable.

Nasal Cannula Inspection: At least once a week, visually inspect
the nasal cannula. Make sure there are no kinks or obstructions.
Replace as needed with a standard 7-foot. (2-meter) single lumen
nasal cannula from your equipment provider.

Reserve Oxygen Supply: Your local equipment provider should
provide or suggest an alternative source of supplemental oxygen
therapy in case there is a mechanical failure or power outage.

Electronic Reset: If the Display Screen does not appear as described
on page 7 or the Reset Required alarm is activated, follow the steps
below to reset the device to default factory settings:

1. Turn the device off.
2. After the device has turned off, press and hold the Power button

until the Greeting Screen stops flashing and the display turns

blank. Once the display is blank, release the Power button.

3. Allow approximately 15 seconds for the device to cycle through a

short sequence. During this time the device will power on and off.

4. Power the device back on. Verify that the Display Screen is

appearing as described on page 7. If this reset does not resolve

the issue, contact your equipment provider.