Inova Labs XYC100B-P4L User Manual

Page 28

Advertising
background image

26

The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-P4L
is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XYC100B-P4L should assure that it is used in
such an environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Immunity Test

IEC 60601

Compliance

Electromagnetic

Test Level

Level

Environment - Guidance

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should

be at least 30%.

+/-6 kV contact

+/-8 kV air

+/-6 kV contact

+/-8 kV air

Electrostatic

Discharge (ESD)

IEC 61000-4-2

Power frequency magnetic

fields should be at levels

characteristic of a typical

commercial or hospital

environment.

3 A/m

3 A/m

Power Frequency

(50/60Hz)

Magnetic Field

IEC 61000-4-8

Main power quality

should be that of a typical

commercial or hospital

environment.

+/-2 kV

for power

supply lines

+/-1kV for input/

output lines

+/-2 kV

for power

supply lines

Not applicable

Electrical Fast

Transient/Burst

IEC 61000-4-4

Main power quality

should be that of a

typical commercial

or hospital

environment.

+/-1 kV

line-to-line

+/-2 kV

line-to-earth

+/-1 kV

line-to-line

+/-2 kV

line-to-earth

Surge

IEC 61000-4-5

Main power quality should be

that of a typical

commercial or hospital

environment. If the user of

the Model XYC100B-

P4L requires continued

operation during power

mains interruptions, it is

recommended that

the Model XYC100B-

P4L be powered from an

uninterruptible power supply

or a battery.

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in U

T

)

for 5 seconds

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in U

T

)

for 5 seconds

Voltage dips,

short interruptions

and voltage variations

on power supply

input lines

IEC 61000-4-11

Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the accompanying tables.

EMC INFORMATION