The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-

P4L

is intended for use in an electromagnetic environment in which radiated

RF disturbances are controlled. The customer or the user of the Model

XYC100B-

P4L

can help prevent electromagnetic interference by maintaining a

minimum distance between portable and mobile RF communications equipment

(transmitters) and the Model XYC100B-

P4L

as recommended below, according to

the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended

separation distance ‘d’ in meters can be estimated using the equation applicable

to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the

transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range

applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation

is affected by absorption and reflection from structures, objects, and people.

Recommended Separation Distances Between Portable and Mobile

RF Communications Equipment and The LifeChoice

®

Activox Portable

Oxygen Concentrator Model XYC100B-P4L

The LifeChoice

®

Activox Portable Oxygen Concentrator Model XYC100B-

P4L

is intended for use in the electromagnetic environment specified below. The

customer or the user of the Model XYC100B-

P4L

should assure that it is used

in such an environment.

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

Separation Distance According to the Frequency of Transmitter (m)

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.2

1.2

2.3

10

3.7

3.7

7.4

100

12

12

23

Rated Maximum

Output Power of

Transmitter (W)

150kHz to 80MHz

80MHz to 800MHz 800MHz to 2.5GHz

d=1.17 √P

d=1.17 √P

d=2.33 √P

Emissions Test

Compliance

Electromagnetic Environment - Guidance

The Model XYC100B-P4L uses RF energy only for

its internal function. Therefore, its RF emissions are

very low and are not likely to cause any interference

to nearby electronic equipment.

RF Emissions, CISPR 11

RF Emissions, CISPR 11

Harmonic Emissions,

IEC 61000-3-2

Voltage Fluctuations/Flicker

Emissions, IEC 61000-3-3

Group 1

Class B

Class A

Complies

The Model XYC100B-P4L is suitable for use

in all establishments, including domestic

establishments and those directly connected to

the public low-voltage power supply network that

supplies buildings used for domestic purposes.