Product information, Inline, Conventional metal-ceramic – Ivoclar Vivadent IPS InLine System User Manual

Page 5: One – one-layer metal-ceramic, Material

Advertising
background image

5

Material

IPS InLine and IPS InLine One are veneering ceramic materials
containing leucite. They are suitable for the fabrication of metal-
ceramic restorations at firing temperatures higher than 900 °C
(1652 °F). With both products, alloys in the CTE range of
13.8–15.0 x 10

-6

/K

-1

(25–500°C) can be veneered, irrespective of the

metal composition. These ceramics are based on leucite-forming
glasses, some of which are produced of feldspar raw materials of a
natural origin. Given their composition they demonstrate excellent
chemical resistance. With the corresponding mixture and targeted
heat treatment of these glasses, leucite crystals with a defined grain
size distribution are released in the glass matrix. This results in a
homogeneous structure for the veneering material, which is not only
extremely gentle to antagonist but also provides the high strength
and convincing optical properties of the IPS InLine veneering ceramic
materials.

Product Information

InLine

®

– conventional metal-ceramic

IPS

InLine

®

One – one-layer metal-ceramic

IPS

Indications

– One-layer veneering ceramic for the most popular dental alloys in

the CTE range of 13.8 – 15.0 x 10

-6

/K (25 – 500°C) (IPS InLine One)

– Conventional multi-layer veneering ceramic for the most popular

dental alloys in the CTE range of 13.8 – 15.0 x 10

-6

/K (25 – 500°C)

(IPS InLine)

– Veneers on refractory die material (only IPS InLine)

contraindications

– If patients are known to be allergic to any of the ingredients, the

material should not be used.

– Bruxism
– Veneering of titanium and zirconium oxide frameworks
– Any other use not listed in the indications

Important processing restrictions

– Exceeding or falling short of the stipulated veneering layer

thicknesses

– Failure to observe the layer thickness ratio between the framework

and layering ceramic

– Mixing with and processing in conjunction with other dental

ceramics

– Veneering of dental alloys not within the stipulated CTE range
– Failure to observe the necessary minimum connector and frame-

work thicknesses

Side effects

If patients are known to be allergic to any of the components in the
materials, IPS InLine One and IPS InLine restorations should not be
used.

5 µm

IPS InLine Deep Dentin, IPS InLine Dentin, IPS InLine Transpa Incisal,
IPS InLine Gingiva, IPS InLine One Dentcisal:

CTE (25 – 500°C) [10

-6

/K]

1)

12.9 ± 0.5

Flexural strength (biaxial) [MPa]

1) 2)

≥ 50

Chem. solubility [μg/cm

2

]

1)

≤ 100

Firing temperature [°C]

900 – 930

1)

according to ISO 6872:2008

2)

typical mean value for the flexural strength is 80 MPa

Classification: Dental ceramics Type 1 / Class 1

Advertising