5.Front Panel Operation

Figure 5-2: Set the frequency using number keys

To Set the Amplitude

Press F2 , confirm whether the "Ampl" menu item is highlighted; if not, press F2 to
switch into "Ampl". In the Parameter bar 2, a cursor appears under the value of
amplitude. Use the knob or the number keys to set the desired value.

To Set the Offset

Press F3 , confirm whether the "Offset" menu item is highlighted; if not, press F3
to switch into "Offset". In the Parameter bar 2, a cursor appears under the value of offset.
Use the knob or the number keys to set the desired value.

To Set the High Level

Press F2 , confirm whether the "Hi_Level" menu item is highlighted; if not, press F2
to switch into "Hi_Level". In the Parameter bar 2, a cursor appears under the value of
amplitude. Use the knob or the number keys to set the desired value.

To Set the Low Level

Press F3 , confirm whether the "Lo_Level" menu item is highlighted; if not, press F3
to switch into "Lo_Level". In the Parameter bar 2, a cursor appears under the value of
offset. Use the knob or the number keys to set the desired value.

To Output Square Signals

Press

button to call the user interface of Square signal, the Square waveform

parameters can be set by operating the Square setting menu on the right.
The parameters of Square waveform are: Frequency/Period, Amplitude/High Level,
Offset/Low Level, Duty. You can operate the menu by using the menu selection buttons
on the right.
To set the Frequency/Period, Amplitude/High Level, Offset/Low Level, please refer to "To

Output Sine Signals" on P11.
The frequency range is 1μHz~50MHz; the period range is 20ns~1Ms.