To set the screen saver, To set output parameter, To set the output load – OWON AG-S series User Manual

Page 36

Advertising
background image

5.Front Panel Operation

(2)

Press F2 to switch between Comma, Space, Off.

Take the Frequency parameter for instance:

Comma

Space

Off

To Set the Screen Saver

The screen saver will run automatically if no operation is taken for any key within the set
time. Press any button to resume.

(1)

Press Utility and choose Disp Setup, press F3 to select Scrn Svr.

(2)

Press F3 to switch between On/Off.

(3)

If On is selected, you can set the screen saver time. Turn the knob to change the
value; or press the number keys to input the desired time in minutes, press F4 to
select the unit. The screen saver time range is 1~999 minutes.

To Set Output Parameter

To Set the Output Load

For the Main output terminal on the Front panel, the Generator has a built-in 50Ωseries
impendence. If the actual load does not match the set one, the displayed amplitude and
offset are incorrect. This function is used to match the displayed voltage with the
expected one.
Steps for setting the Load:

(1)

Press Utility and choose Output Setup, press F1 to select HighZ or *Ω
( "*" represents a value).

(2)

To change the load value, after selecting *Ω, turn the knob to change it directly; or
press the number keys to input the desired value. Press F3 or F4 to select
the unit. The load range is 1Ω~10KΩ.

Note:
The series have

a fixed 50Ω Series Impendence. No matter what Value the set parameter

is, if the real load is different from the set one, the displayed voltage will not equal the real
voltage.