Timer:parameter, Timer:parameter -132 – RIGOL DP832A User Manual
Page 156

RIGOL
2-132
DP800 Programming Guide
:TIMEr:PARAmeter
Syntax
:TIMEr:PARAmeter <secnum>,<volt>,<curr>,<time>
:TIMEr:PARAmeter? <firnum>[,<timercount>]
Description
Set the timer parameters of the specified group.
Query the timer parameters of the specified groups.
Parameter
Name
Type
Range
Default
<secnum>
Integer
0 to 2047
None
<volt>
Real
The voltage range of the
current channel
CH1/CH2: 1V
CH3: -1V
<curr>
Real
The current range of the
current channel
1A
<time>
Real
1s to 99999s
1s
<firnum>
Integer
0 to 2047
None
<timercount>
Integer
1 to 2048
1
Explanation
<secnum> is the group number of the specified group of timer parameters;
<volt>, <curr> and <time> are the voltage, current and time of the specified
group of timer parameters respectively and their units are V, A and s
respectively.
The query command queries the specified groups of timer parameters (the
numbers of the groups are continuous). <firnum> is the group number of the
first group of timer parameters to be queried. <timercount> is the total
number of groups of timer parameters to be queried.
When <timercount> is omitted, the command queries a single group of timer
parameters by default.
command to set or query the timer
parameters.
Return Format
The query returns a string starting with #.
For example, #90000000361,8.000,1.0000,10;2,6.000,1.0000,10;;
wherein, #9000000036 is the data block header;
1,8.000,1.0000,10;2,6.000,1.0000,10; are the specified timer parameters.
The data block header is used to describe the length information of the data
stream and starts with #. For example, the number "9" in #9000000036
denotes that the 9-bit data (000000036) following it is used to denote the data
stream length (36 bytes).
The format of each group of timer parameters is "number,voltage,current,time"
and multiple groups of timer parameters are separated by ";". For example,
1,8.000,1.0000,10;2,6.000,1.0000,10; denotes two groups of timer
parameters; the number of the first group of timer parameters is 1, the voltage
is 8.000V, the current is 1.0000A and the time is 10s; the number of the second
group of timer parameters is 2, the voltage is 6.000V, the current is 1.0000A
and the time is 10s.
Example
:TIME:PARA 1,8,1,10 /*Set the timer parameters of the first group to 8V, 1A, 10s*/
:TIME:PARA 2,6,1,10 /*Set the timer parameters of the second group to 6V, 1A,
10s*/
:TIME:PARA? 1,2 /*Query two groups of timer parameters starting from the
first group and the query returns
#90000000361,8.000,1.0000,10;2,6.000,1.0000,10;*/